Configuration
11
SMA
Nci
3835
8213
ECr r
EH —
Protocol — determines the manufacturer output protocol or serial setting the scale is configured for;
check 3rd party software to confirm correct selection, if the scale is connected to a PC via the
BenchPro
USB port, the USB HID protocol is automatically selected; USB HID settings are
1C19,0002
SMA
— SMA Protocol (all models)Nci — 782X and 76XX family (all models)
3835 — NCI 3835 UPS worldship (models BP-XXXX-XXS only)
8213 — Toledo 8213 (models BP-XXXX-XXS only)
ECr — Electronic Cash Register (models BP-XXXS-XXR only)
EH — Models BP-XXXS-XXS only)
1200
2400
4800
9600
19200
38400
57600
Baud rate — match PC RS-232 setting to the setting of the scale; Check the PC Device Manager
setting through the control panel in port settings
7 even
7 odd
7 none
8 none
Data Bits and Parity — match PC RS-232 setting to the setting of the scale; Check the PC Device
Manager setting through the control panel in port settings
1
2
Stop bits — match PC RS-232 setting to the setting of the scale; Check the PC Device Manager
setting through the control panel in port settings
On
Off
Enable or disable the tare button
No
Yes
Gravity Compensation — see
for detailed information
No - deactivated, calibrate scale with known accurate calibration weights
Yes
- View original calibration gravity and modify local gravity settings
Off
On
Manufacturing mode only; do not use; not not adjust
lb
lb:oz
oz
kg
g
Units of Measure — turn on and off units of measure; Applicable settings are dependant on the
model of scale purchased; see
for available selections; Most models have a
minimum of two units of measure turned on. To avoid incorrect weight being displayed or
transmitted to the PC, only have the applicable units of measure turned on.
150 kg
Capacity — defines the maximum capacity of the scale and determines the weight value to be used;
•
When selecting lb, the calibration weight used must be in lb
•
When selecting kg, the calibration weight used must be in kilograms.
NOTE: See
for capacity selections on each model. Do not select
capacities other than those indicated by the manufacturer,
Off
2rnG
3rnG
2intvl
3intvl
Configure the unit for multi-range or multi-interval on select models of the
BenchPro Series
Off
- unit displays in single range
2rnG - displays weight in multi-range mode dual range
3rnG - displays weight in multi-range mode triple range
2intvl - displays weight in multi-interval mode dual range
3intvl, - displays weight in multi-interval mode triple range
--
Calibration — see
for detailed information on calibration of the scale
RAM
ROM
DIV-A
DIV-O
Diagnostic menu — used to troubleshoot scale operation. See
information on using the diagnostic menu.
RAM - if an error occurs, contact RLWS for a new PCB
ROM - If an error occurs, contact RLWS for a new PCB
DIV-A - Display internal counts after auto zero tracking
DIV-O - Display internal counts
—
Done — exit the configuration menu, save settings and return to weigh mode
Bold
indicates factory default setting.
Parameter
Settings
Description
Table 3-2. Service Settings (Continued)