10
BenchPro Series Operation Manual
4. Press the service setup switch once to enter the service setting mode.
Service
Setup
Switch
5. Set service parameters.
6. Scroll to
once all parameters have been set.
7. Press
to exit and save changes.
Parameter
Settings
Description
2%
5%
10%
20%
Semi Automatic Zero Set — the amount of weight (% of scale capacity) that can be zeroed from the
scale when the zero key is pressed
10%
20%
50%
Initial Power up Zero Setting — the amount of weight (% of scale capacity)that can be zeroed from
the scale upon power up; use if adding a unique weight platter instead of the factory supplied platter
0.25d
0.5d
1d
2d,
3d
Automatic zero tracking — scale returns to zeroweight using the selected multiplier based on the
division or resolution of the displayed weight; in non-approved applications, use this feature to return
the scale to zero weight if in a high vibration environment
Lo
Med
Hi
Filter — controls the weight display update rate; used if the weight displays is fluctuating in a high
vibration environment
Lo - faster update
Med
- default
Hi - slower update
On
Off
Extended Mode — used for engineering testing; turn on this feature to increase the displayed
resolution by one decimal place
Off 1
Off 3
Off 5
Off 30
OFF
Auto Shutdown (battery mode only) — select the amount of time of inactivity after which the scale
automatically powers off
Off 1
— off after 1 minute of no use
Off 3 — off after 3 minuts of no use
Off 5 — off after 5 mintues of no use
Off 30 — off after 30 seconds of no use
Off — scale does not turn off
On
Auto
Off
Backlight Shutdown — conserves battery life; select the amount of time of inactivity after which the
backlight shuts off
On - always on
Auto
- off after 5 seconds no activity
Off - always off
Table 3-2. Service Settings