Request High-Resolution weight (10x)
Command: H (48h)
over capacity (invalid data)
<STX>
?
<S> <CR>
under capacity (-20d)
<STX>
?
<S> <CR>
under zero (Mulis)
<STX>
?
<S> <CR>
not stable
<STX>
?
<S> <CR>
in lb/oz/kg/g (normal data)
<STX> <W>
.
<W> <W> <W> <W> <W> <U> <U> <CR>
in lb/oz/kg/g
<STX> <W> <W>
.
<W> <W> <W> <W> <U> <U> <CR>
in lb/oz/kg/g
<STX> <W> <W> <W>
.
<W> <W> <W> <U> <U> <CR>
in lb/oz/kg/g
<STX> <W> <W> <W> <W>
.
<W> <W> <U> <U> <CR>
in lb/oz/kg/g
<STX> <W> <W> <W> <W> <W>
.
<W> <U> <U> <CR>
in lb:oz
<STX> <W>
l
b
<W> <W>
.
<W> <W> <W>
o
z
<CR>
in lb:oz
<STX> <W> <W>
l
b
<W> <W>
.
<W> <W>
o
z
<CR>
in lb:oz
<STX> <W> <W> <W>
l
b
<W> <W>
.
<W>
o
z
<CR>
in lb:oz
<STX> <sp> <W> <W> <W> <W>
l
b
<W> <W>
o
z
<CR>
Request scale to zero
Command: Z (5ah)
simulate ZERO key
<STX>
?
<S>
<CR>
Scale is placed in echo mode
Command: E (45h)
echo mode enable
<STX>
E
<CR>
Scale is taken out of echo mode
Command: F (46h)
echo mode disable
<STX>
F
<CR>
Scale initiates a test of RAM and ROM
Command: A (41h)
RAM/ROM TEST
<STX>
?
<CR>
Scale confidence test result status
Command: B (42h)
Test result (Command A)
<STX>
<C>
<CR>
Communication