16
BenchPro Series Operation Manual
18. Press
to accept and perform Point1 calibration.
is displayed and then the second linear
calibration weight value.
19. Remove the weight from the weight platter.
20. Press
to accept.
is displayed.
21. Place the weight on the weight platter, wait for the value to stabilize.
22. Press
to accept and perform Point2 calibration. The calibration data is saved and the scale returns to
the weigh mode.
The weight value displayed must match the value of the calibration weight used. If not, perform the calibration a
second time and follow each step carefully. If
displays, there is a calibration error. Ensure the correct
calibration weight value was selected in comparison to the actual calibration weight used.
for more information on troubleshooting.