2.4.4. NIBP messages.
During the normal operation of Omicrom FT monitor, it can show several alarm
messages regarding the NIBP signal (P000PLAC01 board). These messages are:
1) INTERNAL FAILURE
The self calibration has failed or it is not possible to
communicate with the NIBP board.
2) INFLATION TIMEOUT
The time of inflation has exceeded 45 seconds for
adults or 20 seconds for neonates.
3) STEP TIMEOUT
During the measurement the Omicrom FT has been
trying to detect two valid pulses at the same
pressure step during 30 seconds for adults or 20
seconds for neonates.
4) MEASUREMENT
TIMEOUT
The Omicrom FT monitor has been trying to
measure the pressure during more than 2 minutes
for adults or 90 seconds for neonates.
5) WEAK SIGNAL
The pulses of pressure detected at the cuff are too
low and it is not possible to obtain a valid
measurement.
6) AIR LEAKAGE
Three consecutive times the Omicrom FT has
detected a small air leakage in the system, so the
measurement cannot be obtained.
7) DEFLATION FAILURE
During the deflation of the cuff there has been a
failure, so the security valve is opened and the cuff
deflated.
8) OVERPRESSURE
The overpressure detection system has actuated
because the cuff pressure has been higher than the
overpressure limit . The overpressure limit is 315
mmHg for adults and 165 mmHg for neonates.
2.4.5. Invasive pressure messages.
During the normal operation of Omicrom FT monitor, it can show several alarm
messages regarding the IBP signal (P000PLAC01 board). These messages are:
1) INTERNAL FAILURE
The self calibration has failed or it is not possible to
communicate with the IBP board.
2) CABLE OFF
The pressure transducer cable is disconnected from
the IBP board.
3) CHANGE THE SCALE
This message only it is shown when it is selected
the optimum scale level and the pressure signal is
saturated. Select the adequate fix scale or select
Содержание Omicrom FT
Страница 3: ......
Страница 17: ...OMICROM FT SERVICE MANUAL 21 September 1998 2 10 ...
Страница 37: ...OMICROM FT SERVICE MANUAL 21 September 1998 3 10 ...
Страница 54: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 6 5 2 2 P000PLAC01 Part list ...
Страница 57: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 9 5 2 3 P000PLAC04 Part list ...
Страница 59: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 11 5 2 4 P000PLAC05 Part list ...
Страница 61: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 13 5 2 5 P000PLAC07 Part list Component side Solder side ...
Страница 63: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 15 5 2 6 P000PLAC10 Part list ...
Страница 66: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 18 5 2 7 P000PLAC14 Part list ...
Страница 69: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 21 5 2 9 P000PLAC19 Part list ...
Страница 72: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 24 5 2 10 P000PLAC20 Part list ...
Страница 76: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 28 5 3 1 P000PLAC01 Schematics ...
Страница 82: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 34 5 3 2 P000PLAC04 Schematics ...
Страница 85: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 37 5 3 3 P000PLAC05 Schematics ...
Страница 87: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 39 5 3 4 P000PLAC07 Schematics ...
Страница 90: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 42 5 3 5 P000PLAC10 Schematics ...
Страница 95: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 47 5 3 6 P000PLAC14 Schematics ...
Страница 99: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 51 5 3 7 P000PLAC15 Schematics ...
Страница 101: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 53 5 3 8 P000PLAC19 Schematics ...
Страница 108: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 60 5 3 9 P000PLAC20 Schematics ...
Страница 116: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 68 ...