OMICROM FT
SERVICE MANUAL
21 September 1998
2.14
a) Disconnect the cable from the Omicrom FT and check that Omicrom FT
shows an alarm message indicating “Cable off”.
b) Change the invasive pressure alarm levels at Omicrom FT to check that
the patient alarms are being activated correctly.
Pulse oximetry signal verification.
Activate the pulse oximetry signal at the Omicrom FT.
Saturation measurement.
a) Clip the sensor on to the finger.
b) Check that the saturation value showed at Omicrom FT display is
between 92% and 98% (this is a normal value for a healthy adult
person).
c) Check that the heart rate measurement showed at the display is correct.
The actual heart rate can be measured manually over the wrist.
Alarm system verification.
a) Do not clip the sensor on to the finger and check that Omicrom FT
shows an alarm message indicating “Sensor off patient”.
b) Place an opaque means between the emitter and the receiver of the
sensor and check that Omicrom FT shows a message indicating
“Insufficient light”.
c) Change the saturation alarm levels at Omicrom FT to check that the
patient alarms are being activated correctly.
Temperature signal verification.
Connect the temperature simulator (Lionheart 3) to the Omicrom FT using the
special cable supplied with the simulator. To check that the readings of the
Omicrom FT are correct it must be considered the accuracy of the simulator.
Signal activation / disactivation.
a) Connect the temperature probe and check at Omicrom FT display that a
numeric value appears at temperature field.
b) Disconnect the temperature probe and check at Omicrom FT display that
temperature field is blanked.
Temperature measurement.
a) Connect the temperature probe at Omicrom FT.
b) Select at simulator a temperature of 24 ºC and check that the
temperature value showed at Omicrom FT display is correct.
c) Select at simulator a temperature of 37 ºC and check that the
temperature value showed at Omicrom FT display is correct.
d) Select at simulator a temperature of 40 ºC and check that the
temperature value showed at Omicrom FT display is correct.
Temperature alarms:
a) Change the temperature alarm levels at Omicrom FT to check that the
patient alarms are being activated correctly.
Содержание Omicrom FT
Страница 3: ......
Страница 17: ...OMICROM FT SERVICE MANUAL 21 September 1998 2 10 ...
Страница 37: ...OMICROM FT SERVICE MANUAL 21 September 1998 3 10 ...
Страница 54: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 6 5 2 2 P000PLAC01 Part list ...
Страница 57: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 9 5 2 3 P000PLAC04 Part list ...
Страница 59: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 11 5 2 4 P000PLAC05 Part list ...
Страница 61: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 13 5 2 5 P000PLAC07 Part list Component side Solder side ...
Страница 63: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 15 5 2 6 P000PLAC10 Part list ...
Страница 66: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 18 5 2 7 P000PLAC14 Part list ...
Страница 69: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 21 5 2 9 P000PLAC19 Part list ...
Страница 72: ...OMICROM FT SERVICE MANUAL 21 September 1998 5 24 5 2 10 P000PLAC20 Part list ...
Страница 76: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 28 5 3 1 P000PLAC01 Schematics ...
Страница 82: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 34 5 3 2 P000PLAC04 Schematics ...
Страница 85: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 37 5 3 3 P000PLAC05 Schematics ...
Страница 87: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 39 5 3 4 P000PLAC07 Schematics ...
Страница 90: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 42 5 3 5 P000PLAC10 Schematics ...
Страница 95: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 47 5 3 6 P000PLAC14 Schematics ...
Страница 99: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 51 5 3 7 P000PLAC15 Schematics ...
Страница 101: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 53 5 3 8 P000PLAC19 Schematics ...
Страница 108: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 60 5 3 9 P000PLAC20 Schematics ...
Страница 116: ...OMICROM FT 21 September 1998 SERVICE MANUAL 5 68 ...