Error messages and remarks
70
8.2 Remarks
Remarks inform the user about certain device or program processes. The operation of the
device or program may be briefly interrupted but there is no fault. The remark must be
acknowledged by the user to continue the process. Remarks offer the user additional
information but do not represent any device or program faults.
MM 500 cryo
Reference
code
Description
Measures
H10
Motor is overheated
The drive can withstand short-term overloads. In
the case of prolonged overload, self-protection
becomes active. This can happen especially with
high loads (feeding the sample material too
quickly, hard sample).
•
Leave the machine switched on until the cool