Environmental conditions
The following environmental conditions comply with (or exceed) BS EN 61010-1:1993:
Indoor use
IP30 (no protection against water)
Altitude
Up to 2000 m
Operating temperature
0 °C to +50 °C
Storage temperature
‐10 °C to +70 °C
Relative humidity
80% maximum ﴾non‐condensing﴿ for temperatures up to +31 °C
Linear decrease to 50% at +40 °C
Transient overvoltages
Installation category II
Pollution degree
2
NOTE:
If a higher IP rating is required, an additional external enclosure will be required to house the UCC2 and UCC2-2. This
enclosure must facilitate an airflow to allow the internal temperature to maintain the ambient within the operating temperature
range.
UCC2 and UCC2-2 installation guide