20
CMM declutch
"CMM declutch"
21
E-STOP tripped
"ESTOP tripped"
22
Air pressure
Machine air pressure switch
input
"Air pressure switch"
23
Crash
Crash detector switch input
"Crash switch"
24
Contactor feedback
Contactor input signal
"Contactor feedback"
25
X axis outer limit
CMM X axis limit switch input
"Limit switches"
26
-X axis outer limit
CMM X axis limit switch input
"Limit switches"
27
Y axis outer limit
CMM Y axis limit switch input
"Limit switches"
28
-Y axis outer limit
CMM Y axis limit switch input
"Limit switches"
29
Z axis outer limit
CMM Z axis limit switch input
"Limit switches"
30
-Z axis outer limit
CMM Z axis limit switch input
"Limit switches"
31
X axis inner limit
CMM X axis limit switch input
"Limit switches"
32
-X axis inner limit
CMM X axis limit switch input
"Limit switches"
33
Y axis inner limit
CMM Y axis limit switch input
"Limit switches"
34
-Y axis inner limit
CMM Y axis limit switch input
"Limit switches"
35
Z axis inner limit
CMM Z axis limit switch input
"Limit switches"
36
-Z axis inner limit
CMM Z axis limit switch input
"Limit switches"
37
24 V return (0 V)
Common reference line for input
and output signals
38
4th axis brake
Reserved for future use
39
+4th axis outer limit
CMM 4th axis limit switch input
"Limit switches"
40
-4th axis outer limit
CMM 4th axis limit switch input
"Limit switches"
41
+4th axis inner limit
CMM 4th axis limit switch input
"Limit switches"
42
-4th axis inner limit
CMM 4th axis limit switch input
"Limit switches"
43
Reserved
Reserved for future use
"Air pressure switch"
44
24 V return (0 V)
Common reference line for input
and output signals
"24 V return"
Shell
Screen
+24 Vdc
+24 Vdc supply provided for the use of contactors, relays and other motor control circuits. Current limit 1 A.
UCC2 and UCC2-2 installation guide
www.renishaw.com
Issued 07 2021
23