Changing the IP address of the UCC2 or UCC2-2
If the UCC2 or UCC2-2 has an IP address, after booting up it skips its IP configuration state and proceeds into waiting for a download, shown
by the error LED flashing slowly. There are two ways to force the UCC2 or UCC2-2 into its IP configuration state:-
a) During the 15 second power up phase, before the error LED comes on, press and release the Reset button on the rear of the UCC2 or
UCC2-2. At the end of the power up phase the error LED will flash rapidly indicating it is in the IP configuration state.
b) If the UCC2 or UCC2-2 has already powered up with the error light flashing slowly, i.e. about once per second, or even if the control
software has been downloaded, the reset switch will need to be pressed twice. The first time to reset the controller then a second time, within
15 seconds, to force the IP configuration state.
The UCC2 or UCC2-2 is now in its IP configuration state, and the UCC2 or UCC2-2 IP configuration software can be run on the PC, selecting
the appropriate network adapter required. The current IP address of the UCC2 or UCC2-2 will be displayed. Type in the new one and click on
‘Configure PC and UCC pair'.
Establishing a new UCC2 or UCC2-2 PC pairing
If you connect a different UCC2 or UCC2-2 to your PC, or a different PC to your UCC2 or UCC2-2, or even if you modify the network settings
on the PC, you must ensure that they are a matched pair. To do this, go through the procedure in 'Setting the IP address of the PC', even if you
do not wish to change the IP address of the UCC2 or UCc2-2. (If the IP address is already suitable, then you do not need to re-type it.) Clicking
on ‘Configure PC and UCC pair' will establish the new pairing.
UCC2 and UCC2-2 installation guide
www.renishaw.com
Issued 07 2021
38