25
A
– Yes, we can replace the sensor material for
approximately half the cost of a new sensor. Please visit
our website (see cover) for pricing, availability, shipping,
and other information.
Q
– What does “Wheel Slip Error” on the display indicate?
A
– The wheels should be kept dry during operation to
provide adequate traction. They may need to be removed
and cleaned if they have picked up contamination.
Q
– Can the device run across tile grout joints during a
DCOF test?
A
– Most properly installed tile floors will have grouted
joints between flat level tiles, and are traversable, as long
as the device crosses the joints at an angle of
approximately 45 degrees. The test run distance may be
adjusted to fit within most single tiles. Avoid running
across uneven, raised tile edges that protrude above the
normal surface plane, as the sensor could catch on the
raised edge.
Q
– Why should I utilize a Standard Test Method?
A
– Different testing methodologies often produce quite
different results due to the number and degree of
variables. Standard Test Methods are designed to apply
controls and limits, therefore allowing test results to be
used within a context or framework of acceptable and
unacceptable levels of risk.
The current time and date, along with other device settings,
may be edited by navigating to OPTIONS/DEVICE/ DEVICE
SETTINGS. The “SAVE” switch must be pressed when
completed.