22
System
appears
“dead”
Low battery level.
When the level is critically
low, the battery is
automatically disconnected
to protect the battery
Charge the battery.
Battery disconnect
functionality
activated.
Nothing happens.
Activate the battery, see
section 4.2.3.
Battery is worn out.
The battery level indicators
on the hand control drop
quickly after full charge.
Service is needed.
Hand
control
appears
“dead”
Poor contact with
control box.
Nothing happens when
pressing a button, no LEDs
are lit.
Make sure the hand control
is properly connected to the
HC connector.
Low battery level.
Nothing happens when
pressing a button, no LEDs
are lit.
Charge the battery.
Hand control out of
order.
Nothing happens.
Service is needed.
3
Before this happens, the RCB10+ Premium will give a warning sound and the low battery LED on the hand control will flash.