7 Trouble shooting
In addition to this trouble shooting guide, it can be useful to check error and diagnostic flags in the RST10
software, see section 5.3.4.
Problem
Probable cause
Description
Solution
No actuator
movement
Overcurrent
Current peaks above the
current limit are allowed for
250 ms. When this is
exceeded, the actuator will
be turned off.
Overcurrent can be caused
by overload or by running
into an obstacle or a
mechanical end stop.
Overload – reduce load.
Obstacle – remove obstacle.
End stop – run in opposite
direction.
Short circuit
The alert indication LED on
the hand control
is lit.
Service is needed.
Duty cycle protection Duty cycle protection is a
function that protects the
actuator motor from
overheating.
The alert indication LED on
the hand control is lit during
attempt to run the actuator.
Wait until the actuator “rest
time” has elapsed.
Actuator not correctly
connected to control
box
No actuator movement.
Make sure the actuator is
properly connected to an
actuator connector.
Actuator mix up
Wrong actuator will move.
Make sure the
actuator/actuators are
properly connected.
Color coding is possible on
both actuator and control
box.
Internal error
The alert indication LED on
the hand control
is lit.
Wait 10 s. If the error
remains, service is needed.
Actuator cable faulty No actuator movement.
Check the cable and replace
the actuator if necessary.
Charger is connected No actuator movement.
The alert indication LED on
the hand control
is lit during
attempt to run the actuator.
Disconnect the charger.