R&S UPZ
Putting into Operation
Putting into Operation
Operating Manual 1120.8610.14 - 0
5
11
EMI impact on measurement results
Electromagnetic interference (EMI) can affect the measurement results. To avoid any
impact, make sure that the following conditions are met:
●
Use suitable double-shielded cables.
●
Do not use USB connecting cables exceeding 1 m in length.
●
Use only USB devices that remain within the permissible EMI limits.
1.3.1 Unpacking and Checking the Instrument
To remove the instrument from its packaging and check the equipment for
completeness proceed as follows:
1.
Pull off the polyethylene protection pads from the instrument's rear feet and then
carefully remove the pads from the instrument handles at the front.
2.
Pull off the corrugated cardboard cover that protects the rear of the instrument.
3.
Carefully unthread the corrugated cardboard cover at the front that protects the
instrument handles and remove it.
4.
Check the equipment for completeness using the delivery note and the accessory
lists for the various items.
5.
Check the instrument for any damage. If there is damage, immediately contact the
carrier who delivered the instrument. Make sure not to discard the box and packing
material.
Packing material
Retain the original packing material. If the instrument needs to be transported or
shipped at a later date, you can use the material to prevent control elements and
connectors from being damaged.
1.3.2 Placing or Mounting the Instrument
Permissible operating position of the R&S UPZ: horizontal
Operating the instrument in vertical position is not recommended since the narrow
casing (1 HU) may fall over.
For problem-free operation of the instrument, observe the following:
●
Observe the permissible ambient temperature as specified in the data sheet.
●
Avoid condensation. In case condensation has occurred, dry the instrument prior to
switching it on.