The interface is a QSFP+ (Quad Small Form-factor Pluggable) module. It sup-
ports max. bandwidth of up to 50 Gsample/s with optical active cables.
NOTICE!
Risk of damaging the HS DIG I/Q interface.
Use active optical QSFP+ cable with max. power class 2 module (2.0 W max.
power consumption).
●
To plug in the cable:
Hold the QSFP+ plug by its panes. The release tab must be upwards
oriented.
Insert it in the HS Dig I/Q cage and push it.
●
To disconnect the cable:
Do not pull the cable itself.
Puling the cable can damage the cable
and the HS Dig I/Q connector.
Pull the release tab and pull the connector out of the HS Dig I/Q connec-
tor.
See also "Digital Baseband Input Settings" in the user manual.
Adv Data/Ctrl
Interface for exchanging of external data and control signals.
LO In/LO Out
Option: R&S
SMBVB-K90
Local oscillator input and output for phase-coherent RF signal:
●
LO In: Input of phase coherence signal
●
LO Out: Output of phase coherence signal
NOTICE!
Risk of damaging components. Always use an appropriate 8 mm torque
wrench to tighten SMA connectors. Limit the tightening torque to 60 Ncm. Never
use an open-end wrench.
I/Q
I and Q inputs for external analog modulation signal directly fed into the I/Q mod-
ulator.
For more information, see chapter "Overview of the Input and Output Signals and
Connectors" in the user manual.
Ref In/Ref Out
Input/output for external reference signal.
Incl. dedicated connectors for the 1GHz reference signal.
Rear Panel Tour
Содержание SMBV100B
Страница 1: ...R S SMBV100B Vector Signal Generator Getting Started Getting Started Version 02 1423104902 G a2...
Страница 8: ...Contents R S SMBV100B 6 Getting Started 1423 1049 02 02...
Страница 42: ...Instrument Tour R S SMBV100B 40 Getting Started 1423 1049 02 02 Rear Panel Tour...
Страница 66: ...Trying Out the Instrument R S SMBV100B 64 Getting Started 1423 1049 02 02 Generating an EUTRA LTE Signal...
Страница 72: ...System Overview R S SMBV100B 70 Getting Started 1423 1049 02 02 Applications Examples of the R S SMBV100B...
Страница 86: ...Instrument Control R S SMBV100B 84 Getting Started 1423 1049 02 02 Remote Operation over VNC...