EchoView
User’s Guide
19
6
Deploying The EchoView
When deploying a MeshGuard network that utilizes EchoViews,
make sure the rest of the network has been set up before using an
EchoView. The EchoView gets its information about the monitors
on a network from the FMC2000 controller. It does not aggregate
network data from all its nodes. Therefore, make sure the
FMC2000 is operational and in communication with all monitors
on the network, as well as with the EchoView.
Before deploying an EchoView, it is important to have the Pan ID
and Channel number on which the rest of the network is operating.
Without these numbers to program into your EchoView, you
cannot make it communicate with the network.
7
Programming The EchoView
Programming Mode is accessed by first providing the correct password.
Press and hold [MODE] and [N/-] until you see this screen:
Input the 4-digit password:
Increase the number from 0 through 9 by pressing [Y/+].
Decrease the number from 9 through 0 by pressing [N/-].
Step from digit to digit using [MODE].