Accessible settings:
None, V.Fast, Fast, Average, Slow.
Notice:
The higher filtering level the longer stabilization time.
14.3. Autozero
The autozero function has been implemented in order to assure precise
indications. This function controls and corrects
„0”
indication.
While the function is active it compares the results continuously with
constant frequency. If two sequentional results differ less than the declared
value of autozero range, so the scale will be automatically zeroed and the
pictograms
and
will be displayed.
If AUTOZERO is disabled zero is not corrected automatically. However,
in particular cases, this function can disrupt the measurement process e.g.
slow pouring of liquid or powder on the weighing pan. In this case, it is
advisable to disable the autozero function.
Procedure:
•
Enter
<
Weighing>
according to ch. 14 of the manual, select
<
Autozero>
and then set an appropriate value.
Accessible settings:
NO
- Autozero off
YES
- Autozero on
14.4. Minimum weight for different functions (LO)
Parameter
<LO THRESHOLD>
is associated with automatic weighing.
Next weighing will not be saved until the indication goes under the
THRESHOLD LO
(net).
Procedure:
•
After
entering
<
Threshold Lo>
according to ch. 14 of this manual
a keyboard is displayed,