Averaging time
Declaration of process duration in seconds (from
1s to 90s) – measurements carried out in set time
interval are used to calculate mean value, which is
the result of the measurement
Automatic mode
Operation mode, in which the scale automatically
begins the following measuring process, if the load
placed on scale’s weighing platform exceeds value
of mass set in
LO
threshold.
29.3. Carrying out animals weighing procedure
•
Enter the working mode
<
Animals weighing>
according to ch.
29.1 of this user manual,
•
Should the tested animal be weighed in a container, load the container
of scale’s weighing platform and tare its mass,
•
On placing the tested animal on the scale’s weighing platform (in the
tarred container), press
key (Process start), which displays a
message box indicating course of the process,
•
The message box contains data on:
- Progress bar indicated in %,
- Value of averaging time interval, set in loacal parameters,
-
key enabling aborting the process,
•
On completing the measuring process, the value of measured result
is locked in the message box,
•
Confirm completing the process by pressing
key.
30. WORKING MODES - CONTROL OF PACKAGED GOODS
The
<
Control of packaged goods>
working mode proceedes the
control of prepackages in accordance with Welmec regulations. When
the control is started from the scale it is completed after testing the whole
sample (number) of products resulting from the algorithm and depending
on the batch quantity. After the testing has been completed a report is
generated.