Notice:
Users can stop testing at any time by pressing screen button
(stop control) in the bottom part of the display. During the control
procedure other screen buttons, function buttons,
and
are locked.
30.4. Non-destructive control with average tare
If
<
Average Tare Estimation>
has been chosen an average package
weight needs to be estimated by weighing at least 10 packages.
Notice:
In order to test a product in mode „non-destructive control with average
tare”, „S” mass standard deviation of the package estimated form at least
10 packages must not be greater than 0.25 of T error related the average
package weight.
During testing the following data are displayed:
Put empty 1 / 10
- Command concerning the weighing process
and a package number out of the total number
of packages.
X
- Average package weight [g]
0.25T
- Boundary condition in [g]
S
- Standard deviation
Result
- Test result
When the last measurement has been made the following window is
displayed and a report from testing can be saved in the database:
113
Содержание Y/KTP series
Страница 2: ...2 DECEMBER 2011...
Страница 10: ...Y KTP C1 R series main dimensions Y KTP C1 K series main dimensions 5 1 2 Scales of Y KTP D2 series 10...
Страница 150: ...Scale computer cable Scale Kafka printer cable Scale printer CITIZEN EPSON cable Scale I O cable 150...
Страница 175: ...Print the marked report on a conneted printer by pressing Using button view the selected report 175...