35
Analog output must be calibrated (adjusted); using P1 potentiometer set the
correct offset (e.g. for 4 - 20mA output, adjust current so that for 0kg indication
it is precisely 4mA). In case of remaining operation modes proceed
analogously.
10.2.2. Technical Specifications
Operation mode
4 - 20mA, 0 - 20mA, 0 - 10V
Resolution
16 bit
Current output resistance
< 500 Ω
Voltage output resistance
>400Ω
Power supply
24VDC (12 - 30V DC), max 40mA
10.2.3. AN Module: Diagrams of Connection Cables
Voltage output wiring diagram
WIRE COLOURS
Colour
Signal
pink
U
OUT
+
grey
GND
yellow
+24V DC
green
GND
Current output wiring diagram
WIRE COLOURS
Colour
Signal
white
I
OUT
+
grey
GND
yellow
+24V DC
green
GND
10.3. Additional Platform Module, DP6
The DP6 module expands indicator functionality, with it an additional platform
can be operated. The module is intended to be mounted inside the indicator.
For DP6 module (484Rxxxx board), the indicator's housing cover features an
additional cable gland through which an additional platform cable is fed.