8. INSTALLER INSTRUCTION
The PUE HX7 indicator serves as basis of load cell scales.
8.1. Connecting 6-Wire Load Cell
Connect 6-wire load cell to the main board following the diagram below:
6-wire load cell connection
Load cell junction J14
Load cell signals
NOTES
REF+
SENSE +
JP6 not soldered
REF-
SENSE -
JP5 not soldered
IN+
IN-
OUTPUT-
+5V
INPUT+
AGND
INPUT-