9.7. Linearity Correction
9.7.1. Linearity Determination
Declaring mass values for subsequent linearity steps and determining
corrections using scale software.
Procedure:
Enter
<Factory/Section linearity>
submenu and select
<Determine>
parameter. Message
<Continue?>
is displayed.
Press
key to confirm.
<Mass>
edit box is displayed.
Unload the weighing pan.
Enter the required mass value (the first linearity correction point) and
press
key to confirm. Message
<Put weight xxx>
is displayed
(where: xxx – entered mass value) is displayed.
Load the weighing pan remembering that the load mass must equal weight
value specified for the first linearity correction point (it is possible to
preview the weighing result in the displayed window).
Upon result stabilization, press
key to confirm.
<Mass>
edit box is
displayed again (second linearity correction point).
Linearity correction process for the 2nd point is analogous to the linearity
correction process for the 1st point.
Upon declaring requested quantity of linearity correction points, press
key when
<Mass>
edit box is displayed again.
<Section linearity>
submenu is displayed automatically.
Upon pressing
key to confirm zero value for linearity
correction,
message: <Value too low> is displayed.
9.7.2. Corrections
Editing corrections for linearity correction points.
Procedure:
Enter
<Factory/Section linearity>
submenu.
Select
<Corrections>
parameter. A list of determined linearity correction
points is displayed. The list contains previously determined mass values
for each correction point and deviations determined in reading units.
Select requested position,
<Mass>
edit box is displayed.