Load the weighing pan with declared amount of parts. When the indication
is stable (
pictogram is displayed), press
key to confirm the
mass.
•
Single part mass is calculated automatically,
<Parts counting>
working
mode is run and the weighing device displays quantity of single parts
(
pcs
).
Total weight value of all parts loaded onto the weighing pan
cannot be greater than the max capacity value.
Total weight value of all parts loaded onto the weighing pan
cannot be lower than the value determined by „Minimum
reference sample mass” parameter. Unless this condition
is met, the weighing device displays a message: <Sample
mass too low>.
Single part mass must be equal or greater than 0.1 reading
unit. Unless this condition is met, the weighing device
displays a message: <Single part mass too low>.
22.5. Setting Reference Sample Mass by Acquiring Mass of a Single Part
from Database
Upon selecting a product from database, single part mass assigned to the
product (
<Part mass>
entry) is entered automatically.
Procedure:
•
Enter
<Parts counting>
mode and press operation panel key to which
pictogram (product database) is assigned, next select given product
from the list.
Weight value of a single part must be declared for the selected
product. To declare weight value of a single part, enter
products database and edit the selected product.
22.6. Entering Reference Sample Mass to Weighing Device Memory
Reference sample mass of single part can be entered to products database.
Procedure:
•
Press button, select the given product.
•
Determine reference sample mass (read sections 22.3 and 22.4 of this
user manual).