10.2.5. Printout Interval
Parameter allowing you to set frequency of print of
<Weighing printout
template>
when continuous transmission is activated. Interval for printouts
is set in seconds with 0.1 [s] accuracy within 0.1 - 1 000 [s] range.
Procedure:
•
Enter
<Peripherals / Computer / Interval>
submenu,
<Interval>
edit box
is displayed.
•
Enter respective value and press
key to confirm changes.
10.2.6. PUE HX5.EX and E2R System
Establishing connection between the scale and the
E2R System
PC software.
E2R System
is a modular solution designed to comprehensively manage
production processes, some stages of which consist in weighing.
<E2R> can be activated exclusively by an <Administrator>
user. In case of integration of the scale with <E2R System>
software, editing of databases on scales is disabled.
Procedure:
•
Enter
<Peripherals / Computer / E2R>
submenu and set respective
option.
Options:
Connection with
E2R System
inactive.
Connection with
E2R System
active.
10.3. Printer
<Printer>
submenu allows you to:
•
set port for establishing communication with printer,
•
select printer code page,
•
define print start - <Prefix> parameter,
•
define print end - <Suffix> parameter.
To set scale-printer communication parameters go to
<
/ Peripherals /
Printer>
submenu.