User Manual 3162A/B
SCPI Command Reference 4-12
Standard
Waveform
Command
Summary
The Standard Waveform Commands control the various parameters of the
standard output functions. Optional modes are omitted from these commands.
Factory defaults after *RST are shown in bold typeface. Parameter low and
high limits are given where applicable. The Standard Waveforms parameters
could be used for the APPLy command.
Command and Parameters
Low Limit
High Limit
Default
[:SOURce]
:SINusoid:PHASe <value>
0
360
0
:SINusoid:PHASe?
:TRIangle:PHASe <value>
0
360
0
:TRIangle:PHASe?
:SQUare:DCYCle <value>
1
99
50
:SQUare:DCYCle?
:PULSe:DELay <value>
0
99.9
10.0
:PULSe:DELay?
:PULSe:WIDTh <value>
0
99.9
10.0
:PULSe:WIDTh?
:PULSe:TRANsition <value>
0
99.9
10.0
:PULSe:TRANsition?
:PULSe:TRANsition:TRAiling <value>
0
99.9
10.0
:PULSe:TRANsition:TRAiling?
:RAMP:DELay <value>
0
99.9
0.0
:RAMP:DELay?
:RAMP:TRANsition <value>
0
99.9
60.0
:RAMP:TRANsition?
:RAMP:TRANsition:TRAiling <value>
0
99.9
30.0
:RAMP:TRANsition:TRAiling?
:SINC:NCYCle <value>
4
100
10
:SINC:NCYCle?
:GAUSsian:EXPonent <value>
10
100
10
:GAUSsian:EXPonent?
:EXPonential:EXPonent <value>
-100
100
-20
:EXPonential:EXPonent?
:DC <%_amplitude>
-100
100
100
:DC?
Содержание 3162A/B
Страница 4: ...This page was left intentionally blank...
Страница 10: ...3162A B User Manual vi Support Offices 6 2...
Страница 12: ...3162A B User Manual viii This page was left intentionally blank...
Страница 17: ...User Manual 3162A B Getting Started 1 5 Figure 1 2 The Front Panel...
Страница 23: ...User Manual 3162A B Getting Started 1 11 Figure 1 7 WaveCAD Example...
Страница 27: ...User Manual 3162A B Getting Started 1 15 This page was left intentionally blank...
Страница 28: ......
Страница 34: ...3162A B User Manual Configuring The Instrument 2 6 This page was left intentionally blank...
Страница 78: ......
Страница 134: ...User Manual 3162A B Maintenance and Performance Checks 5 32 This page was intentionally left blank...