User Manual 3162A/B
Maintenance and Performance Checks 5-12
Sine Flatness
Level Flatness:
±10% to 10 MHz; ±30% to 200 MHz
Equipment:
Oscilloscope
1.
Configure the Model 3162B as follows:
Function:
Sine wave
Frequency: 10.00 kHz
Amplitude:
3 V
Remote Commands: :RES
:FREQ 10e3
:VOLT 3
:OUTP ON
:INST:SEL B
:VOLT 3
:OUTP ON
2.
Connect the Model 3162B OUTPUT to the oscilloscope, set the
oscilloscope input impedance to 50
Ω
, and set oscilloscope to
display the sinewave within exactly 6 vertical divisions. Repeat this
step for channel B.
3.
Change the Model 3162B FREQ setting to 100 MHz. Verify that
peak-to-peak of the displayed sinewave is within 5.4 to 6.6 divisions.
Repeat this step for channel B.
4.
Change the Model 3162B FREQ setting to 200 MHz. Verify that
peak-to-peak of the displayed sinewave is within 4.2 to 7.8 divisions.
Repeat this step for channel B.
Содержание 3162A/B
Страница 4: ...This page was left intentionally blank...
Страница 10: ...3162A B User Manual vi Support Offices 6 2...
Страница 12: ...3162A B User Manual viii This page was left intentionally blank...
Страница 17: ...User Manual 3162A B Getting Started 1 5 Figure 1 2 The Front Panel...
Страница 23: ...User Manual 3162A B Getting Started 1 11 Figure 1 7 WaveCAD Example...
Страница 27: ...User Manual 3162A B Getting Started 1 15 This page was left intentionally blank...
Страница 28: ......
Страница 34: ...3162A B User Manual Configuring The Instrument 2 6 This page was left intentionally blank...
Страница 78: ......
Страница 134: ...User Manual 3162A B Maintenance and Performance Checks 5 32 This page was intentionally left blank...