User Manual 3162A/B
Maintenance and Performance Checks 5-11
2.
Connect the A channel output to the distortion analyzer input.
3.
Load a 5000-point sine waveform to both A and B channels. Use
TRACe# command or WaveCAD to load the waveform.
4.
Set the distortion analyzer to % distortion measurements, set the
Model 3162B sample clock setting, and verify the distortion reading
for both channels as below.
3162B SCLK Setting Distortion Reading
500e6 <0.5%
50e6
<0.5%
5e6
<0.5%
500e3 <0.5%
50e3
<0.5%
5.
Tune the spectrum analyzer to the carrier frequency and adjust the
gain so the fundamental corresponds to 0 dB.
6.
Connect the Model 3162B output to spectrum analyzer input
through a 20dB feedthrough attenuator. Using the command
FUNC:MODE FIX, modify both channels to output sine waveforms.
7.
Set the Model 3162B FREQ setting and verify the harmonic
distortion levels for both channels as follows:
FREQ Setting Harmonic Level
10 MHz <40 dBc
50 MHz <25 dBc
100 MHz
<25 dBc
200 MHz
<25 dBc
Содержание 3162A/B
Страница 4: ...This page was left intentionally blank...
Страница 10: ...3162A B User Manual vi Support Offices 6 2...
Страница 12: ...3162A B User Manual viii This page was left intentionally blank...
Страница 17: ...User Manual 3162A B Getting Started 1 5 Figure 1 2 The Front Panel...
Страница 23: ...User Manual 3162A B Getting Started 1 11 Figure 1 7 WaveCAD Example...
Страница 27: ...User Manual 3162A B Getting Started 1 15 This page was left intentionally blank...
Страница 28: ......
Страница 34: ...3162A B User Manual Configuring The Instrument 2 6 This page was left intentionally blank...
Страница 78: ......
Страница 134: ...User Manual 3162A B Maintenance and Performance Checks 5 32 This page was intentionally left blank...