User Manual 3162A/B
Maintenance and Performance Checks 5-19
Channel B Adjustment
7.
Change the Model 3162B settings using the following remote
commands:
:FREQ 10 e6.
:INST:SEL B
:FUNC:SHAP SQU
:VOLT 3
:OUTP ON
8.
Connect the B channel output through 20dB attenuator to the
oscilloscope input.
9.
Set oscilloscope input impedance to 50
Ω
. Set the oscilloscope
vertical gain and time base, and adjust C9 on engine board B for best
pulse response.
10. Change the Model 3162B frequency setting to 10 kHz.
11. Remove the 3162B signal from the oscilloscope and adjust the vertical
position to be exactly on the center of the trace.
12. Reconnect the 3162B output to the oscilloscope and adjust R5 on
engine board B for best symmetry around the vertical centerline.