3-4 Maintenance/Troubleshooting
Cardiac Stress Treadmills User Guide
4. Turn both adjustment bolts clockwise 1/4 turn. Test the
belt tension as in Step 2. Repeat if necessary until the belt
runs smoothly without slipping. If more than three
adjustments are necessary, call an authorized technician.
Caution!
Do not overtighten the walking belt. Overtightening
can damage the belt and rollers. Do not torque
adjustment screws beyond 80 in/lb (9 Nm) maximum.
5. Stop the treadmill.
6. Check the position of the rear roller guards and adjust if
necessary (see “Rear Roller Guards” on page 3-5).
Belt Tracking
Perform this procedure whenever the belt moves to one side or
the other.
WARNING! Stay off the belt when adjusting the tracking. Do not
touch or grab the walking belt while it is moving.
1. Use the controls on the monitoring system to start the
treadmill at minimum speed and grade.
2. Increase speed to 6 mph (9.5 km/h) and make the
following adjustment (See Belt Tension for the location
of the bolts):
a. If the belt moves to the right, rotate the right tension
bolt clockwise 1/4 turn.
b. If the belt moves to the left, rotate the left tension
bolt clockwise 1/4 turn.
Adjustment bolts