Cardiac Stress Treadmills User Guide
Receiving and Installation C-3
Installation
If you choose to install your treadmill without the assistance
of your representative, follow the procedure given below:
Tools
•
hex wrench (supplied in kit)
Procedure
1. Ensure that the handrail gaskets have been placed in their
proper location (see drawing below).
2. Insert the front handrail into the holes on each side of the
hood.
3. Use the two hex-head socket screws recessed within the
front of the hood to secure the handrail:
a. Using the hex wrench supplied in the ship kit, turn
the socket screws clockwise.
b. Cover the screws with the caps supplied in the ship
kit.
4. Plug the cable for the optional Emergency Stop Button, if
present, into the connector on the front of the hood
(reference figure above).
♦
If the optional Emergency Stop Button is present, it must remain
connected and secured at all times. It should be disconnected
only by an authorized technician.
Socket
Screw
Cap in
Handrail
Gasket
Emergency Stop
Switch Connector
(optional)
Place
Содержание TM55