C-4 Receiving and Installation
Cardiac Stress Treadmills User Guide
5. (High Voltage units only) Plug the female end of the
power cord into the connector on the front of the
treadmill hood. Be sure the plug is connected securely,
then tighten the clamp using the hex wrench provided.
6. If you want to attach the Q-Stress preamp to the handrail,
follow the installation instructions included with the
preamp.
Connecting to the Monitoring System
1. Be sure the treadmill and the monitoring system are
disconnected from the power source.
2. Inspect the monitoring system interface cable connector,
located on the front of the hood (see figure below). If it
appears damaged, contact an authorized technician
before continuing this installation.
If the connector is undamaged, plug the male end of the
monitoring system interface cable into the connector.
Tighten the connector thumb screws until finger tight.
3. Plug the female end into the treadmill connector on the
back panel of the monitoring system.
4. Verify that the outlet voltage matches the voltage on the
nameplate located on the front of the treadmill hood. If
the voltage is adequate, plug in the power cord. Do not
use this outlet if the voltage is not adequate.