4-128 Menus and Commands
6-00058-05 Rev A
Step
Path:
Main Menu
Service
Diags
Accessor
Step
Use Step to provide low level accessor control. The option
should be used only by trained service personnel.
See Figure 4-192.
Figure 4-192
Step Screen
Parameters
Description
Continue
Y
to continue the test.
N
to cancel the test
If
Y
is selected for the Continue parameter value, the
Response Screen appears. See Figure 4-193.
Figure 4-193
Movement Screen
Parameters
Description
X axis
Displays X axis location
Y axis
Displays Y axis location
Amount
Displays the move step
distance (1, 10–100 mm)
This test should be
used by trained
personnel ONLY!
Continue: N<
Note
There is a .3mm
tolerance for each
x-axis or y-axis
step.The Step
screen display does
not show the exact
step movement.
Use Arrows to Move
X axis : 1057.9mm<
Y axis : 0000.0mm
Amount : 001.0mm
Содержание Scalar 10K
Страница 10: ...1 x Contents 6 00058 05 Rev A...
Страница 22: ...2 xxii Figures 6 00058 05 Rev A...
Страница 24: ...3 xxiv Tables 6 00058 05 Rev A...
Страница 26: ...1 2 About This Guide 6 00058 05 Rev A...
Страница 34: ...1 10 About This Guide 6 00058 05 Rev A...
Страница 36: ...2 2 System Description 6 00058 05 Rev A...
Страница 49: ...Internal Components 2 15 September 2007 Figure 2 7 I E Station Magazines...
Страница 51: ...Connectivity 2 17 September 2007 Figure 2 8 Indirect Fibre Channel Attachment...
Страница 58: ...3 2 System Description 6 00058 05 Rev A...
Страница 252: ...4 174 Menus and Commands 6 00058 05 Rev A...
Страница 254: ...5 2 Processing Media 6 00058 05 Rev A...
Страница 276: ...6 2 Error Messages 6 00058 05 Rev A...
Страница 318: ...6 44 Error Messages 6 00058 05 Rev A...
Страница 324: ...in 6 Index 6 00058 05 Rev A...