5.2.1
Cartridge Present and Clear Sensor Calibration Screen
Figure 5-6 The Calibration/Cartridge Present and Clear Sensor Calibration Screen
This screen shows the value of the emitter intensity determined during calibration of
the optoelectronic pairs for the various sensors. The values on this screen should only
be edited by an authorized service technician.
Information on this screen may be edited by pressing the Edit button. Note that changes
to the screen will not be saved until the Save button is pressed. Pressing the Cancel but-
ton will remove all of the current editing and revert to the last saved version of the
screen.
Section
Field
Description
Option
Calibration
Calibrated
Has the library been successfully calibrated?
Yes or No
Cartridge Present Calibration
Present PWM
Emitter intensity determined during calibration for the opto pair
.
1 to 50
Clear Sensor PWM (Pulse
Width Modulation)
Drive
Emitter intensity determined during calibration for the opto pair.
1 to 50
Left Back
Emitter intensity determined during calibration for the opto pair.
1 to 50
Left Front
Emitter intensity determined during calibration for the opto pair.
1 to 50
Left Side
Emitter intensity determined during calibration for the opto pair.
1 to 50
Right Back
Emitter intensity determined during calibration for the opto pair.
1 to 50
Right Front
Emitter intensity determined during calibration for the opto pair.
1 to 50
Right Side
Emitter intensity determined during calibration for the opto pair.
1 to 50
Table 5-1 Calibration/Cartridge Present and Clear Sensor Calibration Screen Information