Warrior Extreme Blood and Fluid Warmer Instructions for Use
Preparing and Operating the System
QIF-IFU-00001 Rev. A6
Page 9 of 28
Preparing the System for Warming
WARNING:
Follow all the instructions provided by the manufacturer of the
blood/intravenous fluid administration set when administrating blood or fluid through the
Warrior
Extreme.
The Compact Disposable Unit cartridge is provided sterile
in a sterile bag. Visually inspect the Compact Disposable
Unit package before opening to ensure that there is no
damage to the packaging or to the product.
Do not use the Compact Disposable Unit if the sterile
package is damaged or is not fully sealed.
Do not reuse the Compact Disposable Unit. The
Compact Disposable Unit is for a single use only.
YY/MM; Do not use if the date has passed the end of
the indicated expiration month (i.e. 20/03, valid until
31
st
of March 2020).
Do not re-sterilize the Compact Disposable Unit
Insert the battery into the Base Unit and close the side
latches.
Pay attention to the correct battery insertion orientation:
the battery label should be alongside the
Base Unit
label. If
you insert the battery in the wrong orientation, the battery
will not dock with the
Base Unit
. If this happens, remove
the battery, turn it 180 degrees and reinsert into the
Base
Unit
.
For more instructions on how to replace the battery during
administration, see section 6.9.
Open the Compact Disposable Unit sterile package.
Connecting the Base Unit to the Compact Disposable Unit
Note:
When connecting the cable to the Compact
Disposable Unit, verify the protrusion marks on
the connector and on the Compact Disposable