Warrior Extreme Blood and Fluid Warmer Instructions for Use
Specifications and Characteristics
QIF-IFU-00001 Rev. A6
Page 21 of 28
equipment and the other equipment should be observed to verify that they are
operating normally.
Warning: Use of accessories, and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions
or decreased electromagnetic immunity of this equipment and result in improper
operation.
Warning: Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 cm (12
inches) to any part of the Warrior EXTREME, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could
result.
Warning: The Warrior EXTREME needs special precautions regarding EMC and
needs to be installed and put into service according to the specific instructions for
maintaining basic safety and essential performance with regard to electromagnetic
disturbances for the expected service life provided in table 7.
Essential performance of the QiF device is that the outflow temperature stabilizes on
38 ± 2 °C (100.4 ± 3.6 ºF).
QiF device shall only be used with the cables provided by the manufacturer. Base
Unit cable length is 90cm (approximately 34 in.);
Portable and mobile RF communications equipment can affect the QiF device.
The system may recognize the disturbance and initialize a self-test.
Table 7. Electromagnetic interference resistance
Immunity test
IEC 60601
level
Compliance level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD), IEC 61000-4-2
8 kV contact
15 kV air
8 kV contact
15 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 5 %.
Electrical fast
transient/burst,
IEC 61000-4-4
2 kV for power
supply lines
1 kV for
SIP/SOP lines
Not required
The Warrior EXTREME is
battery powered equipment.
Surge, IEC 61000-4-5
1 kV line to line
2 kV line to
earth
Not required
The Warrior EXTREME is
battery powered equipment.