Warrior Extreme Blood and Fluid Warmer Instructions for Use
Warnings
QIF-IFU-00001 Rev. A6
Page 26 of 28
k
The Warrior Extreme requires a fully charged battery for its operation. The battery
does not have a memory effect and may be recharged even though it has not been
fully discharged. If the battery has reached its end of life, please contact the
manufacturer for purchasing a new battery.
11.
Warnings
a
Never throw the Warrior Extreme battery into the trash. The batteries contain toxic
materials and need to be disposed of at a designated battery disposable collection
point.
b
Warming IV fluids/blood at flow rates higher than 200 ml/min (for example by using
a pressure infusion bag or fluid pump) may result in an output temperature lower than
38 ± 2 °C.
c
Do not use the Warrior Extreme in MRI, X-ray and CT environments.
d
Warming medications through the Warrior Extreme was not validated!
e
No modification of this equipment is allowed!
f
Use of the Warrior Extreme not according to its instructions may result in failure of
the system or injury to patient.
g
Do not use the Compact Disposable Unit if the sterile package is damaged.
h
The Compact Disposable Unit is for single use only and does not require any
calibration or maintenance. Do not reuse it! Re-use of the product can cause infection
and contamination.
i
Do not re-sterilize the Compact Disposable Unit