Warrior Extreme Blood and Fluid Warmer Instructions for Use
Preparing and Operating the System
QIF-IFU-00001 Rev. A6
Page 8 of 28
sensors constantly measure the fluid temperature and adjust the power supplied to the
heat exchanger accordingly.
6.
Preparing and Operating the System
Charging the Battery
The Warrior Extreme is powered by a researchable battery, which should be fully
charged before use. The battery shall be charged using the supplied Battery Charger
model FUYUANG FY2552000. Charging requires 120/240 [V], 50/60 Hz.
Plug the Battery into the Battery Charger using the charger
adapter, and plug the Battery Charger into an AC electrical
outlet.
Note:
the battery shall be charged using
only
its
designated charging adapter.
Note:
the battery charger adapter shall be attached to
the battery in the correct orientation: the charger
adapter label in on the
opposite
side of the battery
label, as shown in the picture on the right
Allow the battery to charge until the battery is fully
charged. When the battery is fully charged, the battery
charger green LED lights.
(This is expected to be approximately 4—5 hours if the
battery is empty).
When the battery is fully charged, disconnect the battery
from the battery charger adapter and place the battery
inside the Base Unit or store it for later use.
Note:
The battery has a correct insertion orientation.
Before inserting, make sure the battery label is
alongside the Base Unit label. If the battery is
inserted in the wrong orientation, the battery will
not dock into the Base Unit.
Important!
Recharge the Battery after each use and ensure it is full before each use
Store only fully charged batteries.
When placed in long storage, charge the batteries every 4-6 months.
WARNING:
Never throw the Warrior Extreme
battery into the trash. The battery should
be disposed of at a designated battery disposable collection point.