Warrior Extreme Blood and Fluid Warmer Instructions for Use
QIF-IFU-00001 Rev. A6
Page 18 of 28
Description
LCD Notification
Audio
Notification
User Action (if required)
Fluid is warmer than
41 °C
Fluid is hot!
Tout: XX °C (YY °C)
The LCD blinks
No
''Fluid is Hot'' message might appear
due to flow irregularities (e.g. block,
back flow, large bubbles) or sudden
and significant changes in flow rate
(e.g. changing from 180 ml/min to 40
ml/min).
If none of the above conditions is
apparent, turn off the Base Unit and
replace the Compact Disposable Unit
with a new one.
You can mute the audio notification
with a short press on the Self-
Test/Mute button.
Fluid is warmer than
42 °C
Fluid is hot!!!
Tout: XX °C (YY °C)
The LCD blinks
Steady beep
Fluid outflow
temperature decreased
to 35 °C (95 °F) or
colder, even though it
reached a temperature
above 35 °C (95 °F)
before.
T out indication is 35
°C (95 °F) or below
Short beep
Check if fluid input temperature is at
least 4 °C (39.2 °F) and check the
battery status.
Keep track of the LCD display.
System malfunction
(HW/SW)
System Error
Steady beep Make Sure that the Compact
Disposable Unit is flushed and filled
with fluid and restart the system by
shutting it off and then on again.
If the problem is not resolved shut off
the system, replace the Compact
Disposable Unit with a new one, flush
the air out of the Compact Disposable
Unit, and turn the system back on.
If the problem is not resolved contact
the manufacturer or its representative.
General malfunction *MALFUNCTION!!*
Tout: XX °C (YY °C)
The LCD blinks
Steady beep