Warrior Extreme Blood and Fluid Warmer Instructions for Use
Preparing and Operating the System
QIF-IFU-00001 Rev. A6
Page 15 of 28
WARNING:
Always flush the line with intravenous fluid/blood before administration to
the patient’s body.
Battery Replacement during Administration
The battery status is reported by audio and visual notifications.
Table 1. Battery Indications
Battery Indications
Battery Status
Required Action
Note: Battery is
getting Low!
The LCD blinks, and
a steady beep is
sound.
25—30% capacity
Pay attention, a ''Battery Empty''
message will follow.
Battery Empty
5—10% capacity
The device will continue warming
the fluid until the battery is fully
depleted. The system will then shut
down.
During battery replacement there is no need to stop the fluid flow; although unheated
fluid will flow during this time.
To replace the battery and resume fluid heating, follow the instructions below without
delay.
WARNING:
While the Base Unit is in the “Off”
position the fluid/blood continues flowing through
the system without being warmed until the system is
turned on, completes initialization and starts warming
again.
Shut off the Warrior Extreme Base Unit by pressing the
On/Off switch on the back of the system to “0” position.