QIAsymphony SP/AS Consolidated Operating Guide 12/2017
84
2D bar codes.
If the reagent cartridge is sealed and not pierced, the liquid level of all reagents in the reagent
cartridge is set to the original value. An additional liquid-level check will not be performed.
Both reagent cartridge slots are scanned.
Important
Do not mix enzyme racks, buffer, or magnetic-particle troughs from different
reagent cartridges.
Laser scan — tip rack slots
All 18 tip rack slots are scanned to determine the type of tip rack loaded.
All tip rack slots in which a tip rack was detected are scanned to determine the number of
tips. If a tip is detected in the first and last position of the tip rack, the tip rack will be
categorized as full. If the first or last tip is missing, a full scan will be performed to determine
the number of tips in the tip rack.
Laser scan — unit boxes
The unit box slots are scanned to detect the presence of unit boxes in the 4 slots.
Afterwards, the type (8-Rod Cover or sample prep cartridge) and number of consumables are
determined.
Liquid-level scan of detected reagents
This scan is only performed if the liquid level is not known (e.g., for a partially used reagent
cartridge).
Liquid-level scan of detected reagents.
Liquid-level check of the buffer bottle (if detected).
Liquid-level check of the Accessory Trough (if detected).
Содержание QIAsymphony SP
Страница 213: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 213 This page intentionally left blank...
Страница 214: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 214 This page intentionally left blank...
Страница 216: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 216 www qiagen com...