QIAsymphony SP/AS Consolidated Operating Guide 12/2017
135
Note
: 0 µl is not a valid volume. If a sample position contains no sample volume, clear the sample
assignment from this position (see below).
4.
Press
OK
.
The
Sample Rack Layout
screen appears and the updated volume(s) will be
displayed.
5.
To delete entries for particular sample position(s), select the sample position(s)
and press
Clear
.
Important
If a sample position contains no sample volume, clear the sample assignment
from this position. To do this, select the sample position in the
Sample Rack
Layout
screen, and press
Clear
. When a rack file is used, it is not possible to
clear a sample assignment.
Viewing and editing sample IDs
Samples are automatically assigned default IDs based on their position, slot number, and run ID
(e.g.,
B1_S2_100000061
). Extraction controls are also marked as
EC+
or
EC–
. To view the sample
IDs, press
Zoom In
. Use the arrow buttons to scroll through the sample rack.
If desired, the automatically assigned sample IDs can be edited.
Important
If a rack file was used, the sample IDs cannot be modified.