QIAsymphony SP/AS Consolidated Operating Guide 12/2017
149
Unloading the worktable
After the inventory scan is performed the
Assay Setup/Loading Information
screen appears again.
Proceed as follows:
1.
Open the “Eluate and Reagents” and “Assays” drawers. The
Loading Information
screen
appears.
2.
Press a sample rack to be removed.
The detailed screen for that slot appears.
3.
Unload the selected sample rack from the drawer and then press
Remove
in the touchscreen. If there is a second sample rack, repeat this process for
the other rack.
4.
Press a reagent rack to be removed.
The detailed screen for that slot appears.
5.
Unload the reagent rack from the drawer and then press
Remove
in the
touchscreen. If there is a second reagent rack, repeat this process for the
other rack.
6.
If there is a normalization rack, press this slot.
The detailed screen for that slot appears.
Содержание QIAsymphony SP
Страница 213: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 213 This page intentionally left blank...
Страница 214: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 214 This page intentionally left blank...
Страница 216: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 216 www qiagen com...