QIAsymphony SP/AS Consolidated Operating Guide 12/2017
127
Important
Processed samples will be flagged as “unclear” as soon as the QIAsymphony SP
or QIAsymphony AS is paused and the run is resumed.
Resuming a run
To resume a run, press the
Continue SP
or
Continue AS
button. Processed samples will be flagged
as “unclear” as soon as the QIAsymphony SP/AS is paused and continued.
Stopping a run
If a QIAsymphony SP or QIAsymphony AS run is paused, press the
Stop SP
or
Stop AS
button to
stop the integrated run. On pressing
Stop SP
, all batches currently being processed will be
stopped, although AS batches previously started will be completed. On pressing
Stop AS
, all SP
batches currently being processed will be completed
If the run is stopped, all processed samples are flagged with “invalid”. It is not possible to process
these samples further.
After stopping a QIAsymphony SP or a QIAsymphony AS run or if the run stops due to an error,
the buttons of the affected drawers flash. Press the flashing button(s) to display the warning or
error messages
Содержание QIAsymphony SP
Страница 213: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 213 This page intentionally left blank...
Страница 214: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 214 This page intentionally left blank...
Страница 216: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 216 www qiagen com...