QIAsymphony SP/AS Consolidated Operating Guide 12/2017
190
We recommend using the following formula to calculate the duration of decontamination in
minutes:
Dose (mW x s/cm²) x 10.44 = Duration (seconds)
1.
Remove all removable objects (tubes/plates, adapters, consumables, tip disposal chute)
except for the liquid waste bottle from the drawers.
2.
Enter the
Maintenance
screen, and press
Maintenance SP
or
Maintenance AS
.
The
Maintenance AS
button is only available if you are using
QIAsymphony SP/AS instruments
3.
Press the
Start UV light
button.
The
Input/UV cleanup/Duration
screen will open.
4.
Enter the duration of the decontamination in min.
The default setting is 15 min. The UV irradiation time is dependent on the
pathogen. Use the formula above to calculate the irradiation time and then
enter the time into the input box.
A message appears asking you to check whether all plasticware and
consumables have been removed from the worktable
5.
Confirm that all removable objects have been removed from the worktable
by pressing
OK
.
The UV irradiation procedure starts and the robotic arm moves over the
worktable surface for the set irradiation duration.