QIAsymphony SP/AS Consolidated Operating Guide 12/2017
153
Partial inventory scan
If you need to repeat an inventory scan for the “Eluate and Reagents” drawer (e.g., if a change
has been made on the worktable), you can perform a partial inventory scan. You can choose to
scan the following worktable items separately:
Tip Racks left
Tip Racks right
Adapters left
Adapters right
Reagents LLD
12.5.2
Inventory scan of the “Assays” drawer
The inventory scan of the “Assays” drawer is performed on slots 4–6 as for slots 1–3 of the
“Eluate and Reagents” drawer.
If an inventory scan of the “Assays” drawer needs to be repeated, it is also possible to perform a
partial inventory scan where tip racks and adapters can be scanned separately.
After the inventory scan has been performed, the inventory of the QIAsymphony SP/AS
instruments is updated. The system switches off temporary cooling for the slots and switches on
cooling for loaded slots.
Important
The inventory scan must be performed before a run can be started.
12.5.3
Transfer to a PCR cycler
After assay setup, assays are removed from the QIAsymphony AS and can be manually
transferred to a PCR cycler for detection. A choice of output formats enables use of different PCR
cyclers (e.g., Rotor-Gene Q, 96-well cyclers, 32-capillary cyclers) for detection. Cycler files can
be exported from QIAsymphony SP/AS instruments to selected PCR cyclers.