PUREGAS, LLC
P4200W2 Series Air Dryer User’s Guide
Page 80 of 113
P012308 - Rev. E
9.14.6
Allow the Humidity
reading to rise over 10.0%.
9.14.7
After three (3) minutes,
verify that a Humidity Alarm
appears and the dryer goes
into
SHUTDOWN
mode.
9.14.8
Replace the Humitter into
the Combo Block.
9.14.9
Press the
RESET Button
to clear the alarm.
9.14.10
Reinstall the Front Panel
(section 8.10 ).
NOTE:
If the Humitter is disconnected from the Control Board,
**%
will appear
on the Humidity reading and after 5 minutes the unit will Shutdown. This is to
allow for troubleshooting.
If you are unable to create a Humidity / Shutdown alarm as described, see section
13.11 for troubleshooting information.