PULSAR MEASUREMENT
37
P21 DATEM Distance
Allows the user to determine which echo is to be displayed. On start-up, if
the dBi-M is showing an incorrect reading, then simply enter the distance
from the transducer to the required level and, if an echo is present at this
point, the Gate will establish itself around the chosen echo, DATEM will
update in front of the Gate and reference out any other unwanted echoes.
It should be noted that DATEM will reset to default values whilst performing
this function, and reform itself once it has selected an echo.
Enter distance from the face of transducer to the target in
Measurement
Units
(
P104
).
Progress Bar
The progress bar displays information such as devices found, parameter
changes, and colour the changes to indicate different actions are being
performed.
Receiving and Sending Information
Obtaining an echo for the first time, uploading, or downloading parameters
from the dBi-M, the progress bar will increment in green to signify an action
is being performed:
When the initial trace has been received the PC software will indicate the
information received from the dBi-M in the information box below the
progress bar, such as ‘Get Echo trace’ and ‘Get Datem trace’.
Selecting this icon displays the value for the parameter. If a value has
been sent successfully this should return as 0.
Selecting this icon next to the parameter, enables you to view the options
available, you can choose to select a different value for the parameter or
default it to its factory setting.