AVFM 6.1 INSTRUCTION MANUAL
4
CONNECTIONS
POWER INPUT: 100 to 240 VAC 50/60Hz. No adjustments are necessary for voltages within this range. Connect L
(Live) N (Neutral) and AC Ground.
Optional DC: 9-32 VDC. Connect to + and - terminals.
Optional Thermostat and Heater modules are available rated for 115 VAC or 230 VAC.
IMPORTANT NOTE: To comply with CSA/UL electrical safety standards, AC power input and relay connection wires
must have conduit entry to the instrument enclosure. Installation requires a switch, overcurrent fuse or circuit
breaker in the building (in close proximity to the equipment) that is marked as the disconnect switch.
Risk of electric shock. Loosen cover screw to access connections. Only qualified personnel should access
connections.
Note: Use of instrumentation over 40°C ambient requires special field wiring.
Note: User replaceable fuse is 2 Amp 250V (T2AL250V).
FUNCTION TEST:
Connect the sensor to the sensor terminals as shown on next page, then
apply power. Allow 30 seconds for the AVFM 6.1 to initialize.
A. Place QZ02L sensor (flat to the bottom) in a bucket of water
about 6” deep and press the
programing button to view the
status page which shows level and velocity. Stir the water to see
a velocity reading.