CALIBRATION (cont.)
Level
Max
For QZ02L (submersible, default) level sensor, set the
maximum height the level should reach in the system, in
units configured in the Units/Mode menu. If a Round pipe is
selected in the Channel Setup menu, then this value should
be set to the same value as the inside diameter.
Min
For QZ02L (submersible, default) level sensor, set the
minimum height the level should reach in the system, in
units configured in the Units/Mode menu. Minimum level
for the QZ02L sensor is 1 inch, however, this value could be
set to a larger value if you want to ignore levels greater
than 1 inch.
20mA
Set the 20mA value for the level analog output, in units
configured in the Units/Mode menu.
4mA
Set the 4mA value for the level analog output, in units
configured in the Units/Mode menu.
Lvl Offset
For the QZ02L sensor only. Set to 0.00 when sensor is
mounted at the bottom of a channel/pipe. When the sensor
is mounted above the floor, in order to avoid having the
sensor become coated with debris, enter the distance
between the floor and the bottom of the sensor.
--Calibration-------
Range
Max
180.00 in
Min
8.00 in
Level
Max
180.00 in
Min
1.00 in
20mA
180.00 in
4mA
0.00 in
Lvl Offset
0.00 in
LOE Time
30 sec
Level Damping
Mode
FIR
Percent
10%
Window
1.00 in
Velocity
20mA
10.00 ft/s
4mA
0.00 ft/s
Signal Cutoff
10 %
Velocity Damping
Mode
FIR
Percent
10%
Window
1.00 ft/s
Flow
Min
0.00
/m
USG
20mA
USG
500.00
/m
4mA
0.00 USG /m
Cal Constant
1.000
SENSOR
END VIEW
SIDE VIEW
TOP VIEW
SENSOR
U-SHAPE STAINLESS STEEL
18 GA. RECOMMENDED
HEIGHT AS REQUIRED
USE THE MB-QZ
STAINLESS STEEL
MOUNTING BRACKET
(SUPPLIED)
SENSOR ELEVATED
ABOVE FLOOR OF
PIPE OR CHANNEL
LEVEL
SENSO R