AVFM 6.1 INSTRUCTION MANUAL
26
SPECIAL FUNCTIONS (cont.)
Restore Defaults
Select
US
to erase all user settings and return
the instrument to factory default settings with
US units. Select
Metr
to return the
instrument to factory default settings with
metric units.
New Password
Select any number from
0000
to
9999
and press
. Default setting of 0000 will allow
direct access to the calibration menus. Setting
of any password greater than 0000 will
require the password to be entered to access
the calibration menus.
Press
to return to
Menu Selections
.
SIMULATION
Exercises the 4-20mA (0-5V) outputs, digital display and control relays.
Test
Select
Maximum
and press
to simulate maximum Flow,
Level and Velocity and to output 20mA (5V) to the analog
channels.
Select
Minimum
and press
to simulate minimum Flow,
Level and Velocity and to output 4mA (0V) to the analog
channels.
To simulate an intermediate Flow, Level and Velocity set
Test
to
Actual
and then enter a value for the
Level
and
Velocity
. The Flow calculation, analog
outputs and control relays will respond to the simulated
values.
--Special Functions-
Language
English
Analog Out
4-20mA
Backlight
High
Reset Totalizer
NO
Neg. Totals
NO
Rev. Flow
Off
Capture Par
No
Capture WF
No
Restore Defaults
NO
New Password
0000
--Simulation--------
Test
Actual
Level
4.00in
Velocity
6.33ft/s
Flow
991.41USG/m
4-20mA Level
4.35
4-20mA Velocity14.12
4-20mA Flow
20.00
Relays 1 2 3 4 5 6